Signal Path Compensation
(SPC)
The SPC routine
optimizes the oscilloscope signal path for maximum
measurement accuracy. You can run the routine anytime but you should always
run the routine if the ambient temperature changes by 10 °C (18 °F) or more.
To compensate the signal path, follow these steps:
1. Disconnect
any probes or cables from the channel input connectors.
2. Push the Utility button.
3. Push the System screen button to select Cal.
4. Push the Signal Path screen button.
5. Push OK Compensate Signal Path. This procedure takes several minutes
to complete.
NOTE. The signal path compensation does not include calibration to the probe tip.
Adjust
ing the Oscilloscope
Time and Date
To set your oscilloscope to the current date and time, follow these steps:
1. Push the Utility button.
2. Push the System bottom button to select Cong.
3. Push t
he Set Date & Time bottom button. Use the side menu buttons to
set the date and time values.
4. Push
the OK Enter Date/Time side button to set the oscilloscope date and
time.
4 TDS3000C Series Oscilloscope User Manual