134
To Use
Set up the A Event and B Event from their individual tabs. Then select the A–>B Seq tab and click
Trig After Time. Select the trigger delay time with the Trig Delay control. Use the Horiz Delay
control to delay the acquisition relative to the B event. Adjust the Ref Position and B Trig Level
controls as necessary.
Behavior
Use this trigger sequence to trigger the instrument on the B trigger event after the A event.
When you select Trig After Time, the instrument looks for the trigger conditions specified by the
A event. After the A event arms the trigger system, the instrument delays the trigger for the period
of time specified by the Trig Delay control. After the trigger delay time has expired, the
instrument looks for the B trigger event. When the instrument detects the B event, it delays the
acquisition for the period of time specified by the Horiz Delay.
Trigger on Nth Event (Horizontal Delay Off)
To Use
Set up the A Event and B Event from their individual tabs. Then select the A–>B Seq tab and click
Trig on nth Event. Select the number of B events with the Trig Event control. Adjust the Horiz
Position and B Trig Level controls as necessary.