DC offset zero reset
T
o reset the probe DC offset to the default values, do the following:
1. Connect the probe inputs together with the hook tips.
2. Press and hold the probe BANDWIDTH LIMIT and RANGE buttons until the OVERRANGE LED on the probe begins flashing.
3. When the OVERRANGE LED remains lit (after about 4 seconds), release the buttons.
4. Press the AUDIBLE OVERRANGE button on the probe to store the value. The OVERRANGE LED on the probe turns off to confirm
that the default value for the DC offset has been stored.
5. Repeat steps 2 on page 21 through 4 on page 21 for the other range setting on the probe.
6. Perform the DC Offset Zero procedure as described in the previous section.
Probe operating information
High Voltage Differential Probes THDP0100/0200 and TMDP0200 User Manual 21