Video and General Performance Verication Procedures
10. Set the time bas
e on the second signal to 1 ppm high.
11. Verify that there are no CRC errors on the SDI A input.
12. Record a Pass or F ail as appropriate in the test record.
13. Disconnect the test setup.
HD Pixmon Multiburst
Frequency Response
(Option HD)
This test uses a VM5000 to test the frequency response at the Pixmon output.
1. Set the wav
eform monitor to the factory presets. (See page 60, Restore the
Factory Presets.)
2. Connect t
he TG2000 HDVG1 output to the waveform monitor SDI A input.
3. Select the VM5000 Matrix (1080i 59.94) test signal from the HDVG1. This
signal i
s provided on the User Documentation CD.
4. Press CONFIG > Output > HD Alg PixMon > RGB.
5. Connect the waveform monitor PIXMON output to the VM5000, as described
in the VM5000 manual or online help.
6. Start the VM5000 HD and SD application and set the number of Averages to 4.
7. On the Format tab, select 1080i/60, RGB.
8. On the Measurement tab, select Multiburst and clear any o ther measurement
selections.
9. Run the measurement.
86 Waveform Monitors Specications and Performance Verication

Table of Contents

Other manuals for Tektronix WFM7020

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix WFM7020 and is the answer not in the manual?

Tektronix WFM7020 Specifications

General IconGeneral
BrandTektronix
ModelWFM7020
CategoryMonitor
LanguageEnglish

Related product manuals