E.2.2 Replacing the probe module
1
Press the key on the top of the probe handle and
remove the probe module.
2
Fit a new probe module and engage it in place.
E.3 Regular care
E.3.1 Condensate trap
The fill level of the condensate trap can be read from the markings on the trap. A
warning message is displayed if the level in the condensate trap reaches 90% ( , red
flashing light).
Emptying the condensate trap
The condensate consists of a weak mix of acids. Avoid contact with the skin. Make
sure that the condensate does not run over the housing.
Condensate entering the gas path.
Damage tto tthe ssensors aand fflue ggas ppump!
Do not empty the condensate trap while the flue gas pump is in
operation.
1
Hold the measuring instrument so that the
condensate outlet points up.
2
Open the condensate outlet of the condensate
trap: Push out approx. 7 mm to the stop).
3
Let the condensate run out into a sink .
4
Mop up any remaining drops on the condensate
outlet using a cloth.
5
Close the condensate outlet.
The condensate outlet must be completely closed
(marking), otherwise measuring errors could occur
if external air gets in.