K.6 Recommended rinsing times
Recommended rinsing times in measurements with high concentrations and longer
measurements:
Rinse instrument: Expose probe to fresh air and start flue gas analysis
Parameter Concentration [ppm] Measurement duration [min] Recommended rinsing time [min]
CO 50 60 5
100 30 5
200 20 10
500 10 10
1000 10 15
2000 10 20
4000 5 30
8000 5 60
COlow 10 60 5
20 30 5
50 20 10
100 10 10
200 10 15
500 10 20
NO 50 60 5
100 45 5
200 30 5
500 20 10
1000 10 10
2000 10 20
3000 5 30
NOlow 10 60 5
20 45 5
50 30 5
100 20 10
200 10 10
300 10 20
NO2 10 60 5
20 45 5
50 30 5
100 20 10
200 10 10
500 10 20
SO2 50 60 5
100 30 5
200 20 10
500 15 10
1000 10 10
2000 10 20
5000 5 40
K. Technical data