Parameter Measuring range Accuracy Resolution t90
1
Pabs 600...1150hPa ±10hPa 1hPa -
Temperature (NiCrNi) -40...1200°C ±0.5°C at 0.0...99°C 0.1°C at -40.0...999.9°C depends
±0.5% of readingin rest of range 0.1°C at 1000°C...1200°C on probe
Efficiency 0...120% - 0.1% -
Flue gas loss 0...99,9% - 0,1% -
Flue gas dewpoint 0...99,9°C - 0.1% -
CO2 determination 0...CO2 max. ±0.2 Vol% 0.1 Vol% <40s
(Calculated
from O2)
1 Response time 90%, recommended minimum measurement duration to guarantee correct readings: 3min
For activated single dilution slot 2 (factor 5)
Parameter Measuring range Accuracy Resolution
CO, H2-comp. 700...50000ppm +10% of reading (additional error) 1ppm
COlow, H2-comp. 300...2500ppm +10% of reading (additional error) 0.1ppm
SO2 500...25000ppm +10% of reading (additional error) 1ppm
NO 500...15000ppm +10% of reading (additional error) 1ppm
NOlow 150...1500ppm +10% of reading (additional error) 0.1ppm
With activated dilution of all sensors (optional) (factor 2)
Parameter Measuring range Accuracy Resolution t90
1
O2 0...25Vol.% ±1Vol.% of reading additional error (0...4,99Vol.%) 0.01Vol.% <20s
±0,5Vol.% of reading additional error(5...25Vol.%)
CO, H2-comp. 700...20000ppm +10% of reading (additional error) 1ppm
COlow, H2-comp. 300...1000ppm +10% of reading (additional error) 0.1ppm
NO2 200...1000ppm +10% of reading (additional error) 0.1ppm
SO2 500...10000ppm +10% of reading (additional error) 1ppm
NOlow 150.. .600ppm +10% of reading (additional error) 0.1ppm
NO 500...6000ppm +10% of reading (additional error) 1ppm
1
Response time 90%, recommended minimum measurement duration to guarantee correct readings: 3min
Filter lifetime
Parameter Lifetime
CO, H2-comp. 170000 ppmh
NO 120000 ppmh
K. Technical data
51