EasyManua.ls Logo

THERA-Trainer TIGO - Contraindications and Misuse

Default Icon
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
THERA-Trainer tigo / THERA-Trainer veho - User manual
12
Art. No.: A002-629_Rev 06
Last update: CW19/2018
4.1 Contraindications
Do not use the THERA-Trainer:
users weighing more than - see chapter 19 technical data.
users shorter than 120 cm or taller than 200 cm.
users with serious contractures.
users with ulcers or raw skin that would touch the unit (if in doubt,
seek medical advice).
bone instability / bone metabolism disorders (fractures, osteoporosis,
osteogenesis imperfecta).
in the case of severe fixed contractures / spasticities with lack of flexion /
extension in hip and knee >30°.
unsupervised self-training, if the user cannot operate the THERA-Trainer
without assistance.
acute inflammations - e.g. of the joints.
acute thrombosis.
acute neurological deficiencies.
recently herniated discs.
previously damaged tendons (risk of rupture).
Use THERA-Trainer in consultation with a doctor or therapist:
severe pain during exercise.
dislocation of the shoulder and hip.
COPD (chronic obstructive pulmonary disease).
heart failure.
cardiovascular weakness.
artificial hearts.
Note For users with a short upper body or a low seated height, it might be necessary
to use an appropriate configuration of the THERA-Trainer. Before starting training,
consult a doctor, therapist or trained dealer.
4.2 Misuse
Do not use THERA-Trainer:
Combination with other products emitting ionising radiation (e.g. radiation
therapy, nuclear medicine, etc.)
in rooms with
explosive substances
oxygenated air
The presence of
flammable anaesthetics
volatile solvents
for outdoor exercise.
in ambient conditions not conforming to the requirements.
in a lying position.
in the vicinity of RF surgical equipment and in RF shielded rooms with medical
magnetic resonance imaging systems.

Table of Contents