THERA-Trainer tigo / THERA-Trainer veho - User manual
22
Art. No.: A002-629_Rev 06
Last update: CW19/2018
Transport THERA-Trainer with device base including transport castors as follows:
Tilt THERA-Trainer on device base including transport castors 3.
Pull THERA-Trainer to desired position.
Transport THERA-Trainer with flat device base as follows:
Transport THERA-Trainer with at least 2 persons.
Lift THERA-Trainer by transport strap 1 and under drive unit 2.
Put THERA-Trainer down in desired position.
8.6 Start-up
Starting up the THERA-Trainer:
Connect power plug to socket outlet.
THERA-Trainer boots up. Following the start-up, the THERA-Trainer is ready
for operation.
9 OPERATION
9.1 Exercise preparation
WARNING!
Danger to life due to magnetic field!
Make sure that there is no direct body contact with the THERA-Trainer in the vicinity
of the pacemaker or defibrillator.
Maintain a safety distance of at least 5 cm.
DANGER!
Risk of infection due to contaminated THERA-Trainer!
Wear gloves during cleaning and disinfection.
Observe general precautions to prevent users from being exposed to blood-borne
pathogens.
Wear personal protective equipment (gloves, masks, lab coats) when contact to bodily
fluids such as saliva, blood, semen or vaginal secretion is likely to occur.
No direct patient contact by health care professionals who have weeping wounds
or skin inflammations.
When having direct patient contact as a health care professional who has weeping
wounds or skin inflammations, wear appropriate protective clothing (gloves, scrub
caps, etc.).

Table of Contents