THERA-Trainer tigo / THERA-Trainer veho - User manual
32
Art. No.: A002-629_Rev 06
Last update: CW19/2018
Wristbands for therapy grips
Securing hand using wristbands for therapy grips
Note The wristbands for therapy grips allow paralysed/weakened hands to be fixed
quickly and easily to the therapy grips.
Securing both hands is exclusively possible with the help of an assistant.
Always have an assistant supervise training sessions with two secured hands.
With only one secured hand, unassisted training is possible.
For cleaning the wristbands, see chapter 16 Cleaning and disinfection.
Secure hand(s) as follows:
Put open wristband for therapy grips on table or thigh.
Put hand into wristband for therapy grips.
Thread velcro 2 through bracket and secure wrist.
Put paralysed/weakened hand around therapy grip.
Thread velcro 1 through bracket and secure paralysed/weakened hand
on therapy grip.
9.8 Safety equipment
Voice-controlled emergency stop
Note The voice-controlled emergency stop can be used exclusively in combination
with control and display unit with 5.7", 7" or 10.4" colour screen.
When the arms are fixed in the upper body exerciser so that manual control
is not possible:
Stop THERA-Trainer by giving a loud shout.
Set microphone sensitivity in sub menu Basic settings/Configuration
(see chapter 18.2 Control and display unit with 5.7'' / 7'' / 10.4'' colour screen).
CAUTION!
Risk of injury due to damaged safety equipment.
To protect thumb against injuries:
Before every training, make sure that safety equipment is working correctly.
In case of any malfunction, have the unit repaired by a trained dealer.