33
Art. No.: A002-629_Rev 06
Last update: CW19/2018
THERA-Trainer tigo / THERA-Trainer veho - User manual
Anti-spasm-control
Possible causes for response of anti-spasm-control:
Recognised spasm
Irregular pedalling
Uncontrollable muscular activity
Set motor power exceeded
When anti-spasm-control responds:
THERA-Trainer interrupts power supply to cranks.
Cranks are freely movable.
SPASTICITY or MAX. appears on control and display unit. MOTOR POWER
EXCEEDED or spasm symbol.
After 5 seconds, THERA-Trainer restarts at a speed reduced by 5 rpm.
Set crank direction in sub menu after a spasm was detected (see chapter 18.2
Control and display unit with 5.7'' / 7'' / 10.4'' colour screen or see Control and
display unit tigo/veho 2.7" wireless).
If anti-spasm-control responds repeatedly:
Stop training session.
Restart THERA-Trainer.
Reduce distance between seat and THERA-Trainer.
Reduce crank length (see Foot fixing/leg support).
Reduce number of revolutions.
Increase motor power (see Motor power/joint force).
EMERGENCY STOP button
NoteOnly available on 7" control and display unit.
Press the EMERGENCY STOP button in dangerous situations.
The THERA-Trainer stops immediately and the number of revolutions and
the training resistance slow down to 0.
The control and display unit displays STOP.
To arrange for the EMERGENCY STOP:
Press START button.
- OR -
Pull power plug.
Wait 5 seconds.
Connect power plug.
If the THERA-Trainer immediately starts with the training after you have pressed
the START button:
Press EMERGENCY STOP button.
Contact trained supplier and notify them about the defect.
Do not use the THERA-Trainer for further training.
The THERA-Trainer is in START mode.
Press the START button again to begin the new training session.