53
Art. No.: A002-629_Rev 06
Last update: CW19/2018
THERA-Trainer tigo / THERA-Trainer veho - User manual
6. Up/increase
Move up in direction of arrow.
In menu items where a value is entered (e.g. motor power), the value entered
will increase.
Menu structure control and display unit
1. Level 2. Level Adjustable values,
functions
Factory settings
Neuro Training time 1-180 min 15 min
Direction of rotation
automatic
on/off off
Biofeedback Symmetry bar
Symmetry road
Passive
off
Symmetry bar
(coordination for
upper body training)
Motor power 2-22 Nm
(2-8 Nm for upper
body exerciser)
10 Nm
Spasm detection fine
medium
rough
medium
Spasm program forward
backward
alternating
follow
Forward
Autom. speed adaptation on/off off
Heart rate limit 60-150/min 100/min
Ortho Training time 1-180 min 15 min
Biofeedback
(inactive for upper body
training)
Symmetry bar
Symmetry road
Passive
off
Symmetry bar
Motor power 2-22 Nm
(2-8 Nm for upper
body exerciser)
10 Nm
Muscle support on/off on
Heart rate limit 60-150/min 100/min
Cardio Training time 1-180 min 15 min
Biofeedback
(inactive for upper body
training)
Symmetry bar
Symmetry road
Passive
off
Symmetry bar
Motor power 2-22 Nm
(2-8 Nm for upper
body exerciser)
10 Nm
Heart rate limit 60-150/min 100/min