75
Art. No.: A002-629_Rev 06
Last update: CW19/2018
THERA-Trainer tigo / THERA-Trainer veho - User manual
Voice-controlled
emergency stop
does not respond
Increase microphone sensitivity setting.
Training settings are
incorrect
Set and save basic settings again.
Reset basic settings to factory settings.
Biofeedback symmetry
bar or symmetry road
show incorrect values
(only for control and
display unit with 5.7'', 7"
or 10.4'' colour screen)
Check if crank radius setting is the same on both sides.
Correct if necessary.
Check if user is sitting exactly in the centre (axial position).
Correct if necessary.
Make sure that the user's foot movements are correct:
correct: push
incorrect: pull
If fault occurs repeatedly:
Contact a trained supplier. Have THERA-Trainer recalibrated.
The following appears
on the control and display
unit “SPASTICITY”
Anti-spasm-control has responded.
To continue exercising see chapter 9.8 Safety equipment
The following appears
on the control and display
unit “MOTOR POWER
EXCEEDED”
Anti-spasm-control has responded.
To continue exercising see chapter 9.8 Safety equipment.
Spasm symbol appears
on control and display unit.
Anti-spasm-control has responded.
To continue exercising see chapter 9.8 Safety equipment.
The following appears on
the control and display
unit “WARNING THERA-
Trainer used too heavily.
Reduce pedalling speed”
Reduce pedalling speed.
Make sure that the user does not exceed the upper limits
for number of revolutions (85 rpm) and power (80 Watt).
If the user does not reduce power and number of revolutions,
the error message will appear again.
After three messages, THERA-Trainer will stop the training
session.
The following appears on
the control and display unit
“TRAINING STOPPED
THERA-Trainer used too
heavily”
THERA-Trainer stops the training session due to upper
limits for number of revolutions or power being continuously
exceeded.
“OPEN” is displayed
on the 2.7" control and
display unit:
Locking of the crank is closed.
Open it to perform upper body exercise.
(see chapter 9.6 Locking/unlocking the crank)
The “Unlock” graphic is
displayed on the control
and display unit:
Locking of the crank is closed.
Open it to perform upper body exercise.
(see chapter 9.6 Locking/unlocking the crank)
Error code Troubleshooting