4-13
4.4 Verification when measuring anterior chamber depth and crystal lens thickness
Since there are many waveforms of the cortex etc. of the crystal lens when
measuring an eye with cataracts, the front and rear of the crystal lens may not be
captured.
In this case, adjust the waveform display line on the scanned image up and down,
find and display the position where the largest waveforms are shown, and move
the caliper line to the point of the front of the crystal lens.
Caliper line
Anterior
of cornea
Front of
crystal lens
Waveform that clearly
shows the front of the
cornea
Waveform that
clearly shows the
front of the crystal
lens
Waveform that clearly shows the
rear of the crystal lens
Back of
crystal lens
Waveform of the rear side of
the crystal lens is small.
Waveform of the rear side of
the crystal lens is small.
Waveform of cortex etc.