5-1
5. INSPECTION AND MAINTENANCE
5.1 Warranty
One-Year Limited Warranty
The seller warrants this product to be free from defects in material and
workmanship under the normal use of this product for one (1) year or other term
complying with local regulations from the date of invoice issued by Seller to the
original purchaser.
Lamps, paper and other consumable items shall not be covered by this warranty.
This warranty also shall NOT apply if the product has not been installed, operated
or maintained in accordance with the INSTRUCTION MANUAL of Tomey
Corporation (here in after called “Tomey”). Neither seller not Tomey shall be liable
for any damages caused by purchaser's failure to follow instruction for proper
installation, use and maintenance of product.
This warranty is only applicable to the new product and DOES NOT cover any
damage resulting from or caused by accident or negligence, abuse, misuse,
mishandling, improper modification of this product, by persons other than
personnel duty authorized by Tomey, not to a product whose serial number or
batch number is removed, altered or effaced.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES, EXPRESSED OR IMPLIED (INCLUDING SPECIFICALLY,
WITHOUT LIMITING THE GENERALITY OF THE FOREGOING,
ALL WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE), AND ALL OTHER OBLIGATION AND LIABILITY ON
THE PART OF SELLER AND TOMEY. NEITHER SELLER NOR TOMEY SHALL
BE LIABLE FOR INCIDENTAL,
CONSEQUENTIAL OR SPECIAL DAMAGES UNDER ANY CIRCUMSTANCES
OR FOR MORE THAN REPAIR,
REPLACEMENT OR REFUND OF THE PURCHASE PRICE OF DEFECTIVE
GOODS.
5.2 Operation life
This instrument is designed to have an operation life of 8 years when operated under
the appropriate environment and it is adequately inspected and serviced.
5.3 Inspection
When there is a problem, measurement may not be performed correctly.
Contact Tomey or our local distributor immediately for repair.
When measuring the imitation eyes, check that there is no dust or stain
on them.
To make sure that the internal functions operate correctly and performance is
maintained, use the imitation eyes to check the accuracy before using the instrument.
Inspect method
Use the chin rest pins to fix the imitation eyes to the chin rest. Perform measurement
in the same manner as measuring a patient's eyes.
Check that the measured values are within the range specified on the imitation eyes.