3-53
3) When the “Personal Constant” button (5) is touched, the Personal Value
screen (Fig. 2) appears. Touch the “Print” button (6) to print the personal
constants only. Touch the “Close” button (7) to close the Personal Value
screen and return to the previous screen.
(Fig. 2)
3.6.3. OKULIX
When using the OKULIX calculation result to select IOL, thoroughly
determine the selection by also examining cataract surgery methods,
other inspections and other IOL power calculation formulas. In particular,
in the case of the cornea of a sound eye without having surgery to correct
refractive power of the cornea such as LASIK, carefully compare the
OKULIX calculation result with the result of SRK/T formula etc. and, if
there is a large difference, examine them with due care.
a) Entering optical data on OKULIX screen
When an axial length, anterior chamber depth, lens, and corneal thickness
measurement is completed, the measurement data is entered automatically and
cannot be entered manually. When measurement has not yet been performed,
enter the data as follows.
(Fig. 1)
1) Touch the axial length input field (1), the ACD endothelium input field (2), the
lens input field (3), or the Pachy field (4) to activate it. The keypad appears.
Enter the data and touch the “Enter” key on the keypad to apply the data.
b) Entering ultrasound data on OKULIX screen
When an axial length and anterior chamber depth, and lens measurement is
completed, the data in the Axial, ACD(Epi.), or Lens fields are already entered
automatically and cannot be entered manually. When entering Pachy data
manually, the ACD (Endo.) is automatically calculated.
(6)
(7)
(3) (1)
(2)
(4)