3-43
■
a) Re-measurement of the same patient
■ The deleted data cannot be restored. Carefully check the data before
deleting it.
Retain the patient information and delete only the measurement data for each eye.
Delete all data referring to “3.2.6 Clear all measurement data (preparation for
measuring a new patient)” when the patient is changed.
(Fig. 1)
Touch the “Retake” button (1) for approximately 1 second until a beep sounds to
delete the data. It retains the patient information and the measurement data of the
eye other than the inspected eye being displayed.
3.5.5 Checking the measurement data
a) Intraocular pressure correction
This instrument automatically starts calculation and displays the result when all
items required for calculating the intraocular pressure correction are set.
(Fig. 1)
1) Touch the “IOP” button (1) to open the intraocular pressure correction screen
(Fig. 2). When a measurement has already been completed at this point, the
measured value is displayed in the CCT field (2).
However, if measurement points are set and two or more “CCTs” are included
in them, the average of the CCTs is displayed. When measurement points
are not set, the average of all memory numbers is displayed.
(1)
(1)