3-48
(Fig. 3)
3.6.1 Setting the eye to be measured
TRefer to “ 3.2.7 Selecting the eye ”
3.6.2 IOL power calculation
3.6.2.1 Entering calculation parameters
a) Axial length and anterior chamber depth
When an axial length and anterior chamber depth measurement is completed, the
measurement data is already entered automatically and cannot be entered
manually. (Refer to “3.4.3 e) Selecting specific measurement data to be used for
calculating IOL power.") When measurement has not yet been performed, enter
the data as follows.
(Fig. 1)
1) Touch the axial length input field (1) or the ACD input field (2) to activate it.
The keypad (3) appears. Touch the “Enter” key (4) to apply the data.
[Input range]
Axial length : 13.00 - 45.00 mm
Anterior chamber depth : 0.00 - 10.00 mm
(2)
(1) (2)
(3)
(4)