2.4.3 Screens in each ultrasound mode
a) Axial length measurement screen
(1) Gain display/adjustment field
(2) Mode display – Contact/Immersion
(3) Waveform display area
(4) Corneal gate cursor (only in immersion mode)
The waveform on the right of this cursor position is measured as the waveform of the
cornea.
(5) Lens gate cursor
The waveform between these two cursors is the waveform of the lens.
(6) Retina gate cursor
The waveform on the right of this cursor position is measured as the waveform of the retina.
(7) Gate cursor movement button
Moves the gate cursor.
(8) Level cursor/level line
Measurement data will be taken when the waveform rises above this cursor/line position.
(9) Measurement cursor/measurement line
The distance at this cursor/line should be taken as the measurement data.
(10) Measurement conditions display field