3-23
3.3.5 Browsing corneal thickness measurement value and correcting intraocular pressure
(Fig. 1)
1) Touch the “Pachy -> SNR” button (1) to switch the measurement data list
between Pachy and SNR.
2) Touch the “IOP" button (2) to open the intraocular pressure correction screen
(Fig. 2). When a measurement has already been completed at this point, the
measured value is displayed in the CCT field (3).
3) The input fields of parameter 1 (4) and parameter 2 (5) for the intraocular
pressure correction formula become active when touched, and the keypad
appears. Also, when a formula selection button (6) is touched, the
parameters already set are displayed in the intraocular pressure correction
formula.
[Input range]
Parameter 1 : 0 - 1500
Parameter 2 : 0.00001.0000
4) Enter the intraocular pressure data. The keypad appears when the entry field
(7) is touched.
[Input range]
Intraocular pressure data : 1 - 60.0 (mmHg)
: 1.33 - 79.99 (hPa)
The unit for display can be changed according to “3.9.2c) Settings for various
measurements.”
5) Touch the “Back” button (8) to close the intraocular pressure correction
screen (Fig. 2) and return to the previous screen.
(1)
(2)
(Fig. 2)
(4)
(5)
(6)
(7)
(8)
(3)