5.5 Replacing consumables
5.5.1 Fuses
■ Disconnect the power cord from the outlet when replacing fuses.
Otherwise you may get an electric shock, resulting in death or serious
injuries.
■ Use fuses specifically designed for this instrument.
■ When the instrument does not work correctly after fuses have been
replaced, there may be other causes of the problem. Turn off the
instrument immediately and contact our local distributor.
1) Turn power off and disconnect the power plug from the
outlet.
2) Disconnect the power plug from the power terminal.
3) Insert a flat screwdriver or coin into the slot on the fuse
case located at the bottom of the main unit. Turn the
screw counterclockwise to remove it.
4) Replace the blown fuse with a new one.
5) Install the fuse holder in the reverse order of removal.
5.5.2 Printer paper
■ Always use genuine Tomey paper for the printer. Using other types of
paper may cause printer failure.
■ Do not pull paper forcibly. Trying to pull the paper out may cause printer
failure. Touch the "PRINT" button without any data stored to feed blank
paper.
Replace the paper roll according to the procedure below when red lines appear
on the edges of the printer paper.
1) Push the printer cover notch (1) upward and open it.
2) Remove the old printer paper roll and install a new roll. If
the paper roll is installed in the wrong direction, nothing
will be printed.
3) Close the printer cover with the end of the paper
protruding 2cm from the outlet. Press the cover firmly until
you hear a click.
(Fig. 1)
(1)
(Fig. 1)