3-44
(Fig. 2)
2) The input fields of the parameter 1 (3) and parameter 2 (4) for the intraocular
pressure correction formula become active when touched, and the keypad
appears. Also, when a formula selection button is touched, the already set
parameters are displayed in the intraocular pressure correction formula.
[Input range]
■ Parameter 1 : 0 - 1500
■ Parameter 2 : 0.0000 - 1.0000
3) Enter the intraocular pressure data. The keypad appears when the entry field
(6) is touched.
[Input range]
■ Intraocular pressure data : 1 - 60.0 (mmHg)
: 1.33 - 79.99 (hPa)
4) Touch the “Back” button (7) to close the intraocular pressure correction
screen (Fig. 2) and return to the previous screen.
b) Deleting part of measurement data
(Fig. 1)
1) Touch a memory number button (1) to select the measurement data to be
deleted.
2) Touch the “Delete” button (2) to delete the selected measurement data, and
the average and the standard deviation of the corneal thickness are
recalculated.
3) If you have deleted data mistakenly or want to cancel the deletion, select that
data number and you will see that the “Delete” button (2) will change to the
“Return” button. Touch the “Return” button to cancel the deletion. When new
measurement data is captured, however, the deleted data cannot be restored
by pressing the “Return” button.