3-13
b) Selecting inspection eye
Select the type of eye to be examined according to the patient’s eye.
[Crystal lens]
Phakic eye
Select this in the case of phakic eyes.
Aphakic eye
Select this in the case of aphakic eyes.
IOL (PMMA)
Select this when a PMMA IOL is implanted.
IOL (silicon)
Select this when a silicon IOL is implanted.
IOL (acryl)
Select this when an acrylic IOL is implanted.
User settings
Select this when registering optional materials for IOL eyes.
[Vitreous body]
Vitreous body
Silicon oil
Select an appropriate option according to the material of the vitreous
body.
c) Selecting fitting formula
A fitting formula can be selected.
d) Setting Auto Alignment and Auto Measurement
Use Manual mode only when measurement must be performed even
though the sight-fixing condition is unstable. Correct alignment is
difficult in Manual mode and measurement errors easily occur.
Follow the procedures below to make settings for Auto Alignment and Auto
Measurement.