3-31
(Initial setting and setting range of converted acoustic velocity)
c) Measurement mode
Set the measurement mode.
Auto
Select this for normal measurements.
Auto quick
Select this when measurement is difficult.
Manual
Select this when measurement is difficult in Auto and Auto quick modes.
d) Setting contact/immersion mode
Complete the settings for contact mode and immersion mode referring to
“3.9.2 Measurement setting.”
Phakic eye
Average acoustic velocity for axial length : 1,550m/s, 1,500 - 1,600m/s
Crystal lens acoustic velocity : 1,641m/s, 1,540 - 1,740m/s
Anterior chamber depth acoustic velocity : 1,532 m/s, 1,430 - 1,630m/s
Vitreous acoustic velocity (divisional velocity only) : 1,532 m/s, 800 - 2,000 m/s
Dense cataract eye
Average acoustic velocity for axial length : 1,548m/s, 1,500 - 1,600m/s
Crystal lens acoustic velocity : 1,629m/s, 1,540 - 1,740m/s
Anterior chamber depth acoustic velocity : 1,532 m/s, 1,430 - 1,630m/s
Aphakic eye
Average acoustic velocity for axial length : 1,532 m/s, 1,430 - 1,630m/s
IOL eye
Implanted lens acoustic velocity
(Acryl)
: 2,200m/s, 800 - 3,000m/s
(Silicone)
: 1,049m/s, 800 - 3,000m/s
(PMMA) : 2,718m/s, 800 - 3,000m/s
Anterior chamber depth acoustic velocity : 1,532 m/s, 1,460 - 1,630m/s
Vitreous acoustic velocity : 1,532 m/s, 800 - 2,000 m/s
IOL thickness
(Acryl)
: 0.80 mm, 0.10 - 4.00 mm
(Silicone)
: 1.00 mm, 0.10 - 4.00 mm
(PMMA) : 0.80 mm, 0.10 - 4.00 mm