3-33
b) Setting gate cursor
Settings made here are only effective for the eye currently selected.
Settings cannot be made for both eyes simultaneously. Complete
necessary settings for each eye.
(Fig. 1)
1) Select a gate cursor to be adjusted by touching the gate selection buttons (1).
The selected and active cursor is displayed in red but the other cursors are
displayed in white.
2) Touch the gate cursor movement buttons (2) to set the active gate cursor
position.
c) Re-measurement of the same patient
The deleted data cannot be restored. Carefully check the data before
deleting it.
Retain the patient information and delete only the measurement data for each eye.
Delete all data referring to “3.2.6 Clear all measurement data (preparation for
measuring a new patient)” when the patient is changed.
(Fig. 1)
Touch the “Retake” button (1) for approximately 1 second until a beep sounds to
delete the data. It retains the patient information and the measurement data of the
eye other than the inspected eye being displayed.
(1)
(2)
(1)