3-51
2) You can check the registration content by sliding the IOL data list being
displayed with the scroll bar (3).
3) Select data using the “IOL data selection” buttons (4).
4) Touch the "Edit" button (5) to display the screen for editing the registration
content.
5) Touch the “Apply” button (6) to apply the lens constants and to display them
on the previous screen.
e) Entering parameters for Clinical History Method
Activated only when SRK/T Double K is selected as the formula.
(Fig. 1)
1) Enter the corneal refractive power or radius of corneal curvature before the
surgery for correcting the refractive power in K1pre and K2pre fields (1).
2) Manually enter eye refractive power before refractive correction surgery in
Ref.pre (S,C)(2) and eye refractive power after refractive correction surgery
in Ref.post (S,C) (3).
3) When values are entered in K1Pre, K2Pre, Ref.pre (S,C), and Ref.post (S,C),
the calculation starts automatically and the corneal refractive power or radius
of corneal curvature after the refractive power correction surgery is displayed
in Kpost (4). The data can be entered directly. However, Ref.pre (S,C) and
Ref.post (S,C) cannot be entered.
4) Select the distance between the vertexes (V.D.) from the pull-down menu (5).
C.L./12.0/13.5/14.0/15.5/16.0
(1)
(4)
(5)
(2)
(3)