3.4 Preparation for measurement
3.4.1 Calibration of the Probe
yy
yy
y In case of using the disposal tip probe, be sure to calibrate
it with the disposal tip attached.
yy
yy
y Be sure the tip of the Probe is dry and free from any water,
before calibrating probe.
yy
yy
y In case of using the disposal tip Probe, the disposal tip is
automatically detected by SP-100. Therefore, no measure-
ment can be made without calibrating the Probe.
1) This instrument automatically calibrates the
Probe, when the power is turned on and the
probe is plugged. (Fig. 1)
2) When the probe is removed or the tip of the
probe is removed with disposable tip probe, the
instrument automatically detect the abnormal
situation and shows CALIBRATION screen
(Fig.2).
Check the connection of the probe and tip, then
press "W" for re-calibration.
(Fig. 1)
(Fig. 2)
3.4 Preparation for measurement