3-44
3.10 Battery
3.10.2 Low Battery Warning
1) When capacity of the battery becomes low A,
SP-100 screen starts blinking and beeping (Fig.1).
Under this condition, the unit shuts down after 10
minutes of non-operating time, automatically.
Within 10 minutes, the battery needs to be charged
or measured data will be lost. Auto Power Off
waiting time is 3 minutes, but it will be extended to
10 minutes, when the battery is low, in order to
keep a certain time to start recharging the unit.
When the unit is powered by AC adapter, it starts
charging battery, then blinking and beeping stop
automatically.
2) When the battery is low A, built-in printer be-
comes out of service. When PRINT is pressed, a
warning (Fig.2) comes up on screen. Press F4 key
(B) to go back to main screen.
If the measured data is on screen (Fig.3), when
battery icon becomes to , printout is
available only once within 9 minutes. After 9
minnutes, since the battery becomes low or print-
out is provided once, the printer becomes out of
service and a warning (Fig.2) comes up, when
PRINT is pressed.
(Fig. 1)
(Fig. 2)
(Fig. 3)

Table of Contents

Other manuals for Tomey SP-100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tomey SP-100 and is the answer not in the manual?

Tomey SP-100 Specifications

General IconGeneral
BrandTomey
ModelSP-100
CategoryMedical Equipment
LanguageEnglish

Related product manuals