EasyManuals Logo

Troxler ICO 4740 User Manual

Troxler ICO 4740
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
6-10
If the maximum number of ACFs does not already exist, the oven
displays:
to scroll chars
YES to select char
ENTER when done
Enter the desired ACF ID. Use the numeric keys to enter
numbers and move the cursor to the next position. To enter
alphabetic characters, use the arrow keys to scroll through the
characters. When the desired character is displayed, press
(YES) to move the cursor to the next position. To move the
cursor back one position, press the
(BACK SPACE) key.
NOTE
After the last character has been entered, press (YES)
to accept the character before pressing (ENTER) to
accept the ACF ID.
When the ACF ID is complete, press (ENTER). The oven then
requests the new ACF value:
Input new ACF value
from -5.00 to 5.00%
then press ENTER
The ACF value can be from 5.00 to 5.00%, in increments of
0.01%. Use the numeric keys to enter the ACF. Use the arrow
keys to toggle the sign (positive [+] or negative []) of the
correction factor. When the new ACF value is complete, press
(ENTER). The oven activates the new ACF value, briefly
displays a confirmation message, and returns to the Correction
Menu.
ERASE ACF
To erase one or more ACFs, press (3) at the Correction Menu
shown on page 68. The oven displays:

Table of Contents