ICO 4-15
WARNING! Clean ICO
Exhaust System. See
Operation Manual for
details. Press ENTER
When this warning is displayed, press (ENTER) to
continue with the remainder of this procedure. Upon
completion, clean the exhaust plenum box as
described on page A-5.
14. Press (ENTER) to return to the default screen.
If the Auto-Store option is not enabled, the results
can be stored any time before the next burn by
pressing
(STORE) from the default screen, or by
pressing
(MENU), then pressing (7) (Store). Refer to
the Store Function section on page 5-27 for more
information.
NOTE
A project must be active in order to store. For more
information on using projects, refer to Chapter 7.
If the Auto-Store option is enabled, the oven stores
the results as the next sample number in the active
project file and returns to the default screen. Refer
to the Auto-Store section on page 5-12 for
instructions on enabling this option.
15. Wearing gloves and a face shield, use the basket
carrier to remove the sample from the oven and cover it
with the sample safety cage until the sample has cooled
to room temperature.
16. Determine the aggregate gradation per ASTM standards
C 117 and C 136 (AASHTO standards T11 and T27).