ICO 5-13
NOTE
The Auto-Store sample ID number is not incremented
if sample data is stored manually (see page 5-27).
When using the manual Store function, the operator
enters a sample ID number manually. Therefore,
duplicate sample ID numbers could be stored under
the active project.
For example, if the Auto-Store function is turned on
for a new project, the oven stores the sample data for
the first two burn cycles under sample IDs 1 and 2.
Assume that the Auto-Store is then turned off, and the
operator manually stores data from two more burn
cycles using sample IDs 3 and 4. If the Auto-Store
function is then turned on, the sample data for
successive burn cycles will also be stored using sample
IDs 3, 4, and so on.
To prevent duplication of sample ID numbers, Troxler
recommends either using sample IDs beginning at
100 or higher when storing sample data manually or
always enabling the Auto-Store function.
To access the Auto-Store function, press (4) at the Burn Setup
Menu. The oven displays:
Auto-Store
1. On
2. Off
To return to the Burn Setup Menu without changing the
Auto-Store status, press
(ESC).
To turn the Auto-Store function on, press (1).