Chapter 3
3. Enter the following for LOG 2 and LOG 3 modes:
Select the channels for which you want to log
data. In the case of the DUSTTRAK monitor,
there is only one channel to select: Aerosol.
Start Time
Enter the date and time to begin the sample:
• If you enter a blank for a start date, the
sample begins whenever the specified start
time occurs.
• If you enter a blank for the start time, both
start date and start time are ignored, and
the sample begins when the operator
manually starts the sample.
Enter the log interval to use for the test.
Test length
Enter the length for the sample:
• If you enter a value, the instrument
automatically turns off when the last test is
complete.
• If you enter a blank, the operator must
manually stop the sample.
Enter the number of tests to perform.
tests
If you have specified more than one
of tests, enter the time between tests. If you
enter 0 or blank, the next test is started
immediately after the last test is complete.
While you are entering values for LOG 2 and LOG 3 modes, the
Percent Memory Required is dynamically updated to show the amount
of logger memory required to take the programmed sample. If the
protocol you have defined requires more than 100% of memory, you
can decrease the amount of memory required by manipulating the
following protocol parameters:
• Increase the logging interval.
• Decrease the length for the test.
• Decrease the number of tests.
The settings for each LOG mode must not require more than 100% of
the logger memory.
Note: If the Percent memory required is greater than the Available
memory, the logging instrument automatically stops the test
when memory is full.
4.
When you have finished defining the parameters for LOG 2 and LOG 3
modes, select S
end.