Page 28 of 58 Pages
3. Start/ Stop key
This key commands the following 3 functions:
Starting the process.
Stopping the process.
Canceling the FAIL message from the command panel and
opening the electric door locking if available.
Note:
This key cannot be used during the EXE stage/
Starting the process:
It is active while the autoclave is in standby position, if the door
is closed and water level in the reservoir is normal, pressing this
key starts the selected process.
Stopping the process:
It is active while the autoclave is in process, pressing this key at
any stage of the process stops operation.
Canceling the FAIL message
The end of an aborted process, the FAIL light is turned on and
an error message is displayed on the screen indicating the cause
of the failure.
Pressing this key cancels the displayed message and switches off
the FAIL light.
4. Program key
This key is designed for programming the clock and setting different
parameters by the service technician by means of the UP (5) DOWN
(6) keys.
When the PROGRAM key is pressed, the date is displayed with the
cursor under the day. Pressing the PROGRAM key again moves the
cursor under the month and then on to the year.
After pressing the PROGRAM key again the time of the day will be
displayed with the cursor under the hour. Pressing the PROGRAM
key again moves the cursor to the minutes parameter. Each time the
UP/DN key is pressed, the value of the parameter above the cursor is
changed.
After the date and time parameters are set, pressing the PROGRAM
key shows CODE: 000.
A code known to the technical personnel will be set to change certain
parameters and perform a digital calibration of the system, as
described in detail in the technician section.