APPENDIX 3 ________________________________________________________________________
VAISALA ________________________________________________________________________ 111
RS 232C port of the terminal to connector X17 on the top of the main
board and switch the power on.
When dual loop is used, the jumpers in connector X4 on the module
board must be as shown below.
X4
X4
X1 X2
Set the address of the transmitter, it can be any number
between 1 and 99. In this example the address is 22:
>addr 22
Address : 22
Set the serial bus settings according to your system. This
setting will become valid after next RESET or power off:
>seri 2400 e 7 1 f
2400 E 7 1 FDX
Switch echo on:
>echo on
ECHO : ON
>
Change the serial output mode into POLL:
>smode poll
Serial mode : POLL
NOTE
The SMODE command must be given last.
NOTE
The transmitter outputs no prompt (>) after the SMODE POLL
command and it only reacts to commands which include its address.
Check that the transmitter responds to its address: