CHAPTER 5________________________________________________________________ COMMANDS
VAISALA _________________________________________________________________________ 39
If the settings are correct, press ENT.
If the settings need to be changed, press CL:
the quantity (mA/V) starts flashing; it can be changed with the
arrow keys and acknowledged with the ENT key
the lower limit starts flashing
acknowledge the lower limit with ENT or start changing it by
pressing CL; a new lower limit is given one character at a time
with the arrow keys
the upper limit starts flashing
acknowledge the upper limit with ENT or start changing it by
pressing CL; a new upper limit is given one character at a time
with the arrow keys
When channel 1 has been set, the programme goes on to channel 2;
the procedure is the same as with channel 1.
NOTE
Also the analogue output jumpers must be set to correct places (see
FIGURE 4-2)
Selecting and scaling the analogue output
quantities
Disconnect the security lock jumper!
Select Mode in the main menu and Analog outputs in the Mode
menu: