OPERATING MANUAL_______________________________________________________________
62_____________________________________________________________________M210492EN-A
The stabilization of the sensor can be monitored well by entering c
<cr> instead of the reference value.
Using display/keypad commands
Leave the reference instrument and the transmitter for at least 4
hours in the same space so that their temperatures have time to
equalize. Remove the filter cap prior to calibration.
Disconnect the security lock jumper
before turning the transmitter on!
Check the transmitter against the reference.
Select Cali in the main menu and then T; select two-point
calibration T 2 point cal. Change the first point reading with the
arrow keys and press ENT.
Change the temperature and again check the transmitter against
the reference.
Check that the reading corresponds with the reading of the
reference instrument. If not, adjust the second point.
If necessary, change the second point reading with the arrow keys
and press ENT.
Using LED commands
Leave the reference instrument and the transmitter for at least 4
hours in the same space so that their temperatures have time to
equalize. Remove the filter cap prior to calibration.
Disconnect the security lock jumper
before turning the transmitter on!
Check the transmitter against the reference.
Connect an ammeter/voltmeter to the analogue outputs (connector
X2). Give command ¡¡l¡. At the first calibration point the
LED on the left flashes; adjust the first point (offset) with the
arrow switches to the same reading with the reference and press
ENT switch.
Change the temperature and again check the transmitter against
the reference.