OPERATING MANUAL_______________________________________________________________
76_____________________________________________________________________M210492EN-A
Electromagnetic compatibility
The DMP248 transmitter fulfills the standard EN 61326-1:1997 +Am 1:1998 + Am 2:2001,
Electrical equipment for measurement, control and laboratory use - EMC requirements;
Industrial environment.
Test methods
Emission
Radiated emissions CISPR16 class B (CISPR22 Class B)
Immunity
Electrostatic discharge (ESD) EN/IEC 61000-4-2
EM field EN/IEC 61000-4-3
Industrial environment: 10 V/m
EFT Burst EN/IEC 61000-4-4
Surge EN/IEC 61000-4-5
Conducted RF EN/IEC 61000-4-6
CHAPTER 9 OPTIONS
Power supply 24 VDC (VAC)(standard),
115/230 VAC
Serial interface RS 232C (standard), RS 485/422,
current loop
Cable length 0.56, 2, 5 or 10 metres
Alarm output 2 relays 8 A/ 230 V SPCO (Single Pole
Change Over) for adjustable low and
high alarm
Display cover cover with or without local display & keypad