OPERATING MANUAL_______________________________________________________________
56_____________________________________________________________________M210492EN-A
When LED commands are used and when the two analogue channels
do not output either relative humidity and/or temperature, relative
humidity is calibrated on channel 1 and temperature is calibrated on
channel 2. The calibration ranges are 0...100 %RH and -20...+80 °C.
When the transmitter is calibrated at two points, the points must be
either 50 %RH or 50 C apart from each other.
NOTE
As relative humidity is a temperature dependent parameter, the probe
and the salt bath calibrator have to stabilize to the same temperature
for best accuracy.
Two-point calibration and adjustment adjustment
procedure
A two-point humidity calibrationand adjustment should be performed
in stable conditions using saturated salt solutions as references.
Using serial commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize.
Remove the filter cap on the transmitter.
Disconnect the security lock jumper
before turning the transmitter on!
Place the sensor head in the calibration hole of the LiCl bottle (dry
end reference) in the humidity calibrator.
Wait for 30 minutes.
Give command CRH <cr>, enter the first point value and press
<cr>.
>CRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...
If you want to see how the sensor stabilizes to the humidity in the
calibrator, enter c <cr> instead of the first reference:
RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>