OPERATING MANUAL_______________________________________________________________
60_____________________________________________________________________M210492EN-A
Check the transmitter against the reference.
Give command CT <cr>, enter the first point value and press <cr>:
>CT <cr>
T : xx.x Ref1 ? yy.y <cr>
Press any key when ready
If you want to see how the sensor stabilizes to the reference
temperature, enter c <cr> instead of the first reference:
T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...
After giving the correct temperature value (Ref1) and pressing
<cr> press any key and then <cr>.
Using display/keypad commands
Leave the reference instrument and the transmitter for at least 4
hours in the same space so that their temperatures have time to
equalize. Remove the filter cap prior to calibration.
Disconnect the security lock jumper
before turning the transmitter on!
Check the transmitter against the reference.
Select Cali in the main menu and then T; select one-point
calibration T 1 point cal.
Change the reading with the arrow keys to correspond to the
reference and press ENT.
Using LED commands
Leave the reference instrument and the transmitter for at least 4
hours in the same space so that their temperatures have time to
equalize. Remove the filter cap prior to calibration.
Disconnect the security lock jumper
before turning the transmitter on!
Check the transmitter against the reference.