OPERATING MANUAL_______________________________________________________________
18_____________________________________________________________________M210492EN-A
Sample cell
It may be necessary to use the sample cell if the process (e.g. a pipe) is
too small for the DMP248 sensor head. Furthermore, if the process is
very hot (>80 °C) or particularly dirty, the probe is installed in a
sample cell behind a cooling coil and/or filter. In this case, the
ambient temperature must be at least 10 °C warmer than the process
dewpoint in order to avoid condensation in the sample tubing.
SAMPLE CELL
CLASP NUT.
DMP248SC
THIS EDGE,
INTER AGAINST
PUSH PROBE
TIGHTEN THE
G1/4
S
ISO
228/1
G1/4
S
view
A - A
OUT
IN
6
6
28
68
40
80
Sample gas inlet
use connector
R1/4 ISO 7/1
Sample gas outlet
use connector R1/4 ISO 7/1
G1/2
ISO228/1
FIGURE 3-13 Dimensions (in mm) of the DMP248SC sample cell.
A
25
A
20
fixing screws
(e.g. M6x60 or 1/4" x 11/4")
for DMP248
probe
TOP VIEW SIDE VIEW
22
25
80
40
FIGURE 3-14 Fastening to a metal plate (top view) and on a concrete
frame (side view)