CHAPTER 6______________________________________________ CALIBRATION AND ADJUSTMENT
VAISALA _________________________________________________________________________ 57
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...
Place the sensor head in the calibration hole of the NaCl bottle
(wet end reference) in the humidity calibrator.
Wait for 30 minutes.
Press any key and enter the second point value and press <cr>.
RH : xx.x Ref2 ? yy.y <cr>
The stabilization of the sensor can be monitored here as well by
entering c <cr> instead of the reference value.
Using display/keypad commands
Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize.
Remove the filter cap on the transmitter.
Disconnect the security lock jumper
before turning the transmitter on!
Place the sensor head in the calibration hole of the LiCl bottle (dry
end reference) in the humidity calibrator.
Select Cali in the main menu and then RH; select Not changed
and then two-point calibration RH 2 point cal. Change the first
point reading with the arrow keys to correspond the reference
humidity and press ENT; pressing an arrow once changes the
reading by 0.05 %RH.
Place the sensor head in the calibration hole of the NaCl bottle
(wet end reference) in the humidity calibrator.
Wait for 30 minutes.