EasyManua.ls Logo

Vaisala DMP248 - Maintenance; Reference measurements; Self-diagnostics

Vaisala DMP248
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 6______________________________________________ CALIBRATION AND ADJUSTMENT
VAISALA _________________________________________________________________________ 63
Check that the reading corresponds with the reading of the
reference instrument. If not, adjust the second point.
If necessary, adjust with the arrow switches to the correct value
and press ENT. At the second calibration point the second LED
from the left flashes.
Calibration of the analogue outputs
The analogue outputs have been calibrated at the factory and since
they are very stable, calibration of the outputs should be performed
only when there is reason to believe that their adjustments have
changed.
Using serial commands
Disconnect the security lock jumper!
ACAL <cr>
The outputs on channels 1 and 2 are measured and the measured
values (mA or V) entered as calibration coefficients.
Example: both channels have 0...10 V outputs (set with AMODE
command); enter the voltages measured at the analogue outputs:
>ACAL <cr>
Ch1 U1 ( V ) ? 0.123 <cr>
Ch1 U2 ( V ) ? 9.98 <cr>
Ch2 U1 ( V ) ? 0.120 <cr>
Ch2 U2 ( V ) ? 9.98 <cr>
Using display/keypad commands
Disconnect the security lock jumper!
Connect an ammeter/voltmeter to the output of channel 1, select
Cali in the main menu and Analog outputs in the Cali menu. The
following is displayed (the quantity can be either mA or V):

Table of Contents

Other manuals for Vaisala DMP248

Related product manuals