CHAPTER 5________________________________________________________________ COMMANDS
VAISALA _________________________________________________________________________ 51
Stopping the measurement output
S<cr>
Ends the RUN mode; after this command all other commands can be
used.
Outputting the reading once
SEND <cr> in STOP mode
or
SEND aa <cr> in POLL state
aa = address of the transmitter when more than one
transmitter is connected to a serial bus (0...99)
Output:
Td= -9.3 'C PPM= 2733 T= 22.1 'C RH= 10.4 %RH
Td= -9.3 'C PPM= 2730 T= 22.1 'C RH= 10.4 %RH
...
The output format can be changed with command FORM (see
Appendix 1).
Setting the output interval for the RUN mode
INTV xxx yyy <cr>
xxx = output interval (0...255)
0: no pause between outputs
yyy = unit (s, min or h)
Example: output interval is changed into 10 minutes
>INTV 10 min <cr>
Output intrv. : 10 min