•  even or odd parity, 8 data bits, 2 stop bits: if this combination is
given the programme changes the number of stop bits to 1
NOTE
The serial bus settings become effective only after reset.
Setting the transmitter address
Address is used when more than one transmitter is connected to one
serial bus; it makes it possible to communicate with one transmitter at
a time.
•  Select More in the main menu and Addr in the More menu; the
following is displayed:
•  Pressing ENT returns the programme to the main menu.
•  Pressing CL deletes the old address; enter the new address with
the arrow keys.
Selecting the output units
•  Select Unit in the main menu:
•  Use the arrow keys to select the right alternative and press ENT.
metric non-
metric
RH %RH %RH
T °C °F
Td °C °F
ppm
v
ppm ppm