USER'S GUIDE ____________________________________________________________________
56 __________________________________________________________________ M210543EN-F
Automatic Message Sending
In the automatic (CLOSE) mode PWD22/52 sends the predefined
message at defined intervals.
The AMES command defines the automatic message internal AMES
Message_number Message_interval
where
Message_number
=
Valid range 0 ... 7. Selects the corresponding
message.
Negative message numbers are
converted to 0. If only the message number is
given, the previous message interval setting is
used.
Message_interval
=
Given in multiples of 15 s (= measuring
interval).
Therefore, intervals 15, 30, 45... are
valid. Other intervals are converted to
multiples of 15 s.
The maximum sending
interval is 255 s (4 min 15 s).
See the example below:
AMES 0 60
Selects message number 0 to be sent once in a minute.
Messages can also be displayed in the command mode with the MES
command, as described in section MES on page 60.