20 EX10xxA Introduction
GENERAL SPECIFICATIONS
*Note, self-cal is not applicable for RX1032
VOLTAGE OFFSET STABILITY
±10.0 V
±1.0 V
±0.1 V
±0.067 V
±0.01 V
± 20 µV/°C typical
± 10 µV/°C typical
± 5 µV/°C typical
± 2 µV/°C typical
± 2 µV/°C typical
VOLTAGE GAIN STABILITY
Voltage input channels (all ranges)
and thermocouple input channels
*Note, self-cal is not applicable for RX1032
± 25 ppm/°C without self-cal* (typical)
± 5 ppm/°C with self-cal* at any operating temperature (typical)
INPUT IMPEDANCE
40 MΩ differential
INPUT BIAS CURRENT
5 nA typical
COMMON MODE INPUT RANGE
±10 V
COMMON MODE REJECTION RATIO (CMRR)
4 Hz filter
1 kHz filter
dc: 100 dB minimum; (50/60) Hz: 140 dB typical, 120 dB minimum
dc: 100 dB minimum; (50/60) Hz: 100 dB typical, 80 dB minimum
See Figure 7-2 in the Theory of Operation for more information regarding CMRR.
INPUT PROTECTION
±15 V
NETWORK CONNECTION
10/100 Base-T
INPUT CONNECTORS
Thermocouple
EX10xxA
RX10xx
Analog voltage
EX10xxA
EX1000A-51C
EX1000A-TC
Cu-Cu mini-TC jack
Screw terminals (three terminals for each channels “S, +, -“) M3 X 6MM (PAN HEAD PHILLIPS)
50-pin female D-sub
50-pin female D-sub
Cu-Cu mini-TC jack
OPERATING TEMPERATURE (FOR EX10XX A)
0 °C to 50 °C
POWER INPUT (EX10XXA, 51C)
EX10xxA
RX10xx
(90 264) V ac, (50/60) Hz
INPUT VOLTAGE DC, 18-36V DC
DIMENSIONS ( EX10XXA/RX10XX )
EX10xxA
EX1000A-51C
RX10xx
1.75’’ H x 17.5’’ W x 13.6’’D
1.75’’ H x 17.5’’ W x 13.6’’D
4.92” H x 9.05” W x 9.45” D; 125 x 238 x 249 mm (including handles and latches)
WEIGHT (RX10XX )
RX10xx
5.3 KG; 11 lbs
RELIABILITY (EX10XXA)
MTBF
Methodology
> 100,000 hrs (or approximately 11 years)
Telcordia (Bellcore) SR-332, Issue 1
LXI SPECIFICATIONS
LXI CLASS COMPLIANCE
LXI Class A
CLOCK SPECIFICATIONS
Clock oscillator accuracy
Synchronization accuracy
Timestamp
Accuracy
Resolution
±50 ppm
Reports “synchronized” when < ±200 µs of the 1588 master clock
As good as time synchronization down to 50 ns
25 ns
IEEE 1588-BASED TRIGGER TIMING
Alarm