www.vtiinstruments.com
EX10xxA Introduction 21
Trigger time accuracy
Time to trigger delay
Receive LAN[0-7] Event
Trigger time accuracy
Time to trigger delay
Future timestamp
Past/zero timestamp
As good as time synchronization down to 50 us
50 us
As good as time synchronization down to 50 us
50 us typical
1 ms maximum
LXI Trigger Bus
Time to trigger delay
DIO Bus
Time to trigger delay
55 us typical
57 us typical
ENVIRONMENTAL SPECIFICATIONS FOR EX10XXA, EX1000A-51C
This chassis should be operated indoors in a controlled environment, protected from
exposure to the elements (i.e. direct sunlight, precipitation, wind, etc.).
0 C to +50 C
-40 C to +70 C
5% – 95% (non-condensing)
Conforms to MIL-PRF-28800F, Paragraphs 4.5.5.3.1 (random vibration test), 4.5.5.3.2
(sinusoidal vibration test), and 4.5.5.4.1 (functional shock test)
ENVIRONMENTAL SPECIFICATIONS FOR RX10XX
-40°C TO +71°C (REDUCED ACCURACY), 0°C TO +50°C (FULL ACCURACY)
IP 65 RATED ENCLOSURE, 5% TO 95%
CONFORMS TO MIL-PRF-28800F
RANDOM VIBRATION 30 MIN PER AXIS, 10-500Hz MIL-PRF-28800F CLASS 2
SINUSOIDAL 5 TO 55HZ RESONANCE SEARCH PER MIL-PRF-28800F CLASS 2,
EACH AXIS SHOCK 30G/AXIS, 11MS HALF SINE PULSE PER MIL-PRF-28800F
CLASS 2
6 X M6 X 10mm LG, CAP FLANGE HD, 10.9 STEEL SCREWS
ENVIRONMENTAL SPECIFICATIONS FOR EX1044
-10°C TO +45°C (REDUCED ACCURACY), 0°C TO +50°C (FULL ACCURACY)
5% – 95% (non-condensing)
Conforms to MIL-PRF-28800F, Paragraphs 4.5.5.3.1 (random vibration test), 4.5.5.3.2
(sinusoidal vibration test), and 4.5.5.4.1 (functional shock test)