Conditions
• 60 minute warm-up
• Guaranteed maximum limits are two times (2x) the typical values
• 7 days, ±5 °C from last self-calibration***
• 20 °C to 30 °C, 1 year from full calibration
• Exclusive of thermocouple errors
• Exclusive of noise
• Common mode voltage = 0
* 1400 accuracy is for 1372 °C
** 500 accuracy is for 400 °C
Note: Typical is based on 2σ & Max is based on 4σ
Conditions
• After 1 hour warmup period and temperature stable for 1 hour after any temperature change.
• Accuracy does not include accuracy of thermocouple or thermocouple wiring.
• Maximum Temperature rate of change between any operating temperature is 0.5 °C/min
• Guaranteed maximum limits are two times (2x) the typical values
• 7 days, ±5 °C from last calibration
• -20 °C to +30 °C, 1 year from full calibration
• Exclusive of thermocouple errors
• Exclusive of noise
• Common mode voltage = 0
• Self calibration is not applicable
Temperature rate of change
• Nominal - 0.33 °C /min
• Max - 1.33 °C /min