EasyManua.ls Logo

VTI Instruments EX1000A - Page 22

VTI Instruments EX1000A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VTI Instruments Corp.
22 EX10xxA Introduction
TABLE 1-2: THERMOCOUPLE ACCURACY FOR EX10XXA (TYPICAL)
Type
Min
Max
-100
0
100
300
500
700
900
1100
1400
J
-200
1200
±0.25
±0.20
±0.20
±0.25
±0.30
±0.30
±0.35
±0.45
K*
-200
1372
±0.25
±0.20
±0.20
±0.20
±0.35
±0.35
±0.45
±0.55
±0.50
T**
-200
400
±0.25
±0.20
±0.20
±0.20
±0.25
E
-200
900
±0.25
±0.20
±0.20
±0.20
±0.25
±0.30
±0.35
S
-50
1768
±1.00
±0.75
±0.65
±0.65
±0.65
±0.70
±0.70
±0.75
R
-50
1768
±1.00
±0.75
±0.60
±0.60
±0.60
±0.60
±0.65
±0.70
B
250
1820
±1.65
±1.10
±0.80
±0.70
±0.65
±0.65
N
-200
1300
±0.40
±0.25
±0.25
±0.25
±0.30
±0.35
±0.40
±0.40
Values in °C
Conditions
• 60 minute warm-up
• Guaranteed maximum limits are two times (2x) the typical values
• 7 days, ±5 °C from last self-calibration***
• 20 °C to 30 °C, 1 year from full calibration
• Exclusive of thermocouple errors
• Exclusive of noise
• Common mode voltage = 0
* 1400 accuracy is for 1372 °C
** 500 accuracy is for 400 °C
TABLE 1-3: THERMOCOUPLE ACCURACY FOR RX1032
-40 °C to +50 °C
+50 °C to +70 °C
+15 °C to +25 °C
Type E, J, K, N & T
Typical
±0.22°C
±0.53°C
±0.12°C
Max
±0.35°C
±0.93°C
±0.20°C
Type R & S
Typical
±0.88°C
±2.12°C
±0.48°C
Max
±1.40°C
±3.72°C
±0.80°C
Type B
Typical
±1.76°C
±4.24°C
±0.96°C
Max
±2.80°C
±7.44°C
±1.60°C
Note: Typical is based on 2σ & Max is based on
Conditions
• After 1 hour warmup period and temperature stable for 1 hour after any temperature change.
• Accuracy does not include accuracy of thermocouple or thermocouple wiring.
• Maximum Temperature rate of change between any operating temperature is 0.5 °C/min
• Guaranteed maximum limits are two times (2x) the typical values
• 7 days, ±5 °C from last calibration
-20 °C to +30 °C, 1 year from full calibration
• Exclusive of thermocouple errors
• Exclusive of noise
• Common mode voltage = 0
Self calibration is not applicable
Temperature rate of change
Nominal - 0.33 °C /min
Max - 1.33 °C /min

Table of Contents

Related product manuals