EasyManua.ls Logo

VTI Instruments EX1000A - Page 277

VTI Instruments EX1000A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
www.vtiinstruments.com
EX10xxA Function Calls 277
Error Code
Error Message
Meaning
0xBFFC0902L
#define VTEX10XXA_ERROR_INVALID_TRIG_INF_VALUE
Invalid trigger infinite value.
0xBFFC0903L
#define VTEX10XXA_ERROR_INVALID_ARM_INF_VALUE
Invalid arm infinite value.
0xBFFC0904L
#define VTEX10XXA_ERROR_INVALID_NUM_SCANS
Invalid number of scans.
0xBFFC0905L
#define VTEX10XXA_ERROR_READ_FIFO_TIMEOUT
Timeout on read FIFO.
0xBFFC0906L
#define VTEX10XXA_ERROR_STRMING_DATA_ENABLED
Streaming data already enabled.
0xBFFC0907L
#define VTEX10XXA_ERROR_ENABLING_STRMING_DATA
Enabling streaming data failed.
0xBFFC0908L
#define VTEX10XXA_ERROR_STRMING_DATA_NOT_ENABLED
Streaming data not enabled.
0xBFFC0909L
#define VTEX10XXA_ERROR_DISABLING_STRMING_DATA
Disabling streaming data failed.
0xBFFC090AL
#define VTEX10XXA_ERROR_INSUFFICIENT_FIRMWARE_REV
Insufficient firmware revision.
0xBFFC090BL
#define VTEX10XXA_ERROR_VXI_11
VXI-11 Error.
0xBFFC090CL
#define VTEX10XXA_ERROR_INVALID_FILTER_FREQ
Invalid filter frequency.
0xBFFC090DL
#define VTEX10XXA_ERROR_STRMIN_ALREADY_ENABLED
Streaming data already enabled on instrument.
0xBFFC090EL
#define VTEX10XXA_ERROR_INVALID_NUM_ENTRIES
Invalid number of entries.
0xBFFC090FL
#define VTEX10XXA_ERROR_NEGATIVE_TIME
Illegal negative time.
0xBFFC0910L
#define VTEX10XXA_ERROR_DEVICE_LOCKED
Device locked.
0xBFFC0911L
#define VTEX10XXA_ERROR_DEVICE_NOT_LOCKED
Device not locked.
0xBFFC0912L
#define VTEX10XXA_ERROR_CALIBRATION_IN_PROGRESS
Calibration in progress.
0xBFFC0913L
#define VTEX10XXA_ERROR_CALIBRATION_MISSING
Calibration missing.
0xBFFC0914L
#define VTEX10XXA_ERROR_TRIGGER_SYSTEM_ERROR
Trigger system error.
0xBFFC0915L
#define VTEX10XXA_ERROR_INSUFFICIENT_UPTIME
Insufficient uptime for requested operation.
0xBFFC0916L
#define VTEX10XXA_ERROR_OUT_OF_LINK_IDS
Maximum number of link IDs exceeded.
0xBFFC0917L
#define VTEX10XXA_ERROR_INVALID_LINK_ID
Invalid link ID.
0xBFFC0918L
#define VTEX10XXA_ERROR_DUPLICATE_SCANLIST_CHNL
Duplicate channel In scanlist.
0xBFFC0919L
#define VTEX10XXA_ERROR_SELF_CAL_MISSING
Self-calibration missing.
0xBFFC091AL
#define VTEX10XXA_ERROR_UNKNOWN_FIRMWARE_ERROR
Unknown firmware error.
0xBFFC091BL
#define VTEX10XXA_ERROR_INVALID_DIO_LIMIT
Invalid DIO limit value.
0xBFFC091CL
#define VTEX10XXA_BOOLEAN_OUTA_RANGE
Out of range Boolean value.
0xBFFC091DL
#define VTEX10XXA_NOT_A_VALID_RSRC
Not a valid resource string.
0xBFFC091EL
#define VTEX10XXA_ERROR_ALREADY_LOCKED_U
Instrument already locked by this host.
0xBFFC091FL
#define VTEX10XXA_ERROR_ALREADY_LOCKED_NU
Instrument already locked by another host.
0xBFFC0920L
#define VTEX10XXA_ERROR_ALREADY_UNLOCKED
Instrument is not locked at this time.
0xBFFC0921L
#define VTEX10XXA_ERROR_NONVOL_SELF_CAL_MISSING
Nonvolatile self calibration data missing.
0xBFFC0922L
#define VTEX10XXA_ERROR_FIFO_EMPTY
FIFO is empty.
0xBFFC0923L
#define VTEX10XXA_ERROR_STRING_TOO_BIG
Input buffer contains too many characters.
0xBFFC0924L
#define VTEX10XXA_ERROR_PTP_NOT_RUNNING
Precision time protocol not running. Try
reconfiguring time settings, or a power cycle.
0xBFFC0925L
#define VTEX10XXA_ERROR_INVALID_BOOLEAN_VALUE
Invalid Boolean value.
0xBFFC0926L
#define VTEX10XXA_ERROR_WIREDOR_MUST_BE_SET
Wired-OR must be set.
0xBFFC0927L
#define VTEX10XXA_ERROR_INVALID_TIMER_COUNT
Invalid timer count.
0xBFFC0928L
#define VTEX10XXA_ERROR_INVALID_TIMER_TIME
Invalid time setting.
0xBFFC0929L
#define VTEX10XXA_ERROR_INVALID_LAN_SOURCE
Invalid LAN source.
0xBFFC092AL
#define VTEX10XXA_ERROR_INVALID_LAN_EVENT_DOMAIN
Invalid LAN domain.
0xBFFC092BL
#define VTEX10XXA_ERROR_INVALID_LAN_EVENT_EVENTID
Invalid LAN Event ID.
0xBFFC092CL
#define VTEX10XXA_ERROR_INVALID_LAN_EVENT_FILTER
Invalid LAN filter.
0xBFFC092DL
#define VTEX10XXA_ERROR_INVALID_OP_WHILE_LOG_ENABLED
This operation is not valid while LAN event
logging is enabled.
0xBFFC092EL
#define VTEX10XXA_ERROR_PPS_REQUIRES_DIO0_OUTPUT_LOW
PPS enable requires DIO0 output to be enabled
and low.
0xBFFC092FL
#define VTEX10XXA_ERROR_GAIN_OUT_OF_RANGE
Linear correction gain value out of range.
0xBFFC0930L
#define VTEX10XXA_ERROR_OFFSET_OUT_OF_RANGE
Linear correction offset value out of range.
0xBFFC0931L
#define VTEX10XXA_ERROR_INVALID_CORRECTION_LIST_LENGTH
Linear correction list length must be the same as
channel list length.
0xBFFC0932L
#define VTEX10XXA_ERROR_INVALID_GAIN_LIST_LENGTH
Gain list length must be the same as channel list
length.
0xBFFC0933L
#define VTEX10XXA_ERROR_OPERATION_VALID_ONLY_ON_VOLT_CH
This operation is only valid on general purpose
voltage channels.
0xBFFC0934L
#define VTEX10XXA_ERROR_INVALID_EVENTLOG_OVERFLOWMODE
Invalid event log overflow mode.
0xBFFC0935L
#define VTEX10XXA_ERROR_INVALID_CAL_FILE_TYPE
Invalid calibration file type.
0xBFFC0936L
#define VTEX10XXA_ERROR_ACQUISITION_IN_PROGRESS
Acquisition in progress.
0xBFFC0937L
#define VTEX10XXA_ERROR_FACTORY_MODE_ONLY
Only works when calFileType parameter is a
value greater than 0.
0xBFFC0938L
#define VTEX10XXA_ERROR_CAL_FILE_TOO_LARGE
Unable to load the entire calibration file into
memory.