Setting Initial Operating Parameters 191
6. Allow sufficient time for the sample to clear the injector before repositioning the
Load/Inject handle. Reposition the Load/Inject handle as follows:
• If you are operating in the gradient mode, or if the next sample is not ready to be
loaded, leave the handle in the Inject position.
• If you are operating in the isocratic mode, or if the next sample is ready to be
loaded, turn the Load/Inject handle clockwise to the Load position.
11.1.2 Using a Non-PowerLine Autoinjector
For complete information on your non-PowerLine autoinjector, see the operator’s manual
included with the injector.
Injecting Sample
1. Ensure that you have connected the injector according to the instructions in the
Waters 600E Multisolvent Delivery System Installation and Maintenance Guide (see
Section 3.5.2, Connecting a Non-IEEE-488 Autosampler).
2. Load the sample vials into the autoinjector.
3. Program the autoinjector as necessary.
4. Start the autoinjector.
11.2 Setting Initial Operating Parameters
This section describes how to set the initial operating parameters for running the gradient
and event tables that you program (see Chapter 10, Creating Gradient Controller
Time-Based Tables).